User Contributed MET/CAL PROCEDURE ============================================================================== INSTRUMENT: DATA TECH 320 SERIES DATE: 22-Apr-87 AUTHOR: User Contributed REVISION: V 4.0 ADJUSTMENT THRESHOLD: 90% NUMBER OF TESTS: 15 NUMBER OF LINES: 60 CONFIGURATION: Fluke 5100B Fluke 8506A CONFIGURATION: ============================================================================== STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK+ R N M A L T W 1.002 DISP Allow a minumum warm-up of one hour. 1.002 DISP NOTE: GUARD TERMINAL SHOULD ALWAYS BE CONNECTED 1.002 DISP TO GROUND. 1.002 DISP Set UUT to DC VOLTS on the 10 VDC range. 1.002 DISP Short all three terminals together and adjust front 1.002 DISP panel zero control for a reading of 1.002 DISP 00.00 +/- .01 1.002 DISP With terminals still shorted set range to 1000 VDC 1.003 DISP And adjust R358 for a reading of 000.0 +/-.1 1.003 DISP Remove short and connect test cable to terminals 1.004 DISP Remember to connect GND to LO. 1.005 HEAD DC VOLTAGE TESTS 1.006 5100 10 10V 0.02% 2W 2.001 5100 10 -10V 0.02% 2W 3.001 5100 100 100V 0.02% E 2W 4.001 5100 1000 500V 0.02% 2W 5.001 OPBR Does the UUT have Ohms function? 5.002 JMPF 10.001 5.003 HEAD RESISTANCE TESTS 5.004 DISP Set UUT to 100Kohms range with all three input 5.004 DISP terminals shorted together. 5.004 DISP Adjust R67 for a reading of 000.00 +/-.01 5.005 DISP DO NOT USE GUARD IN OHMS TESTS. 5.006 RESF A 10MZ 0.02% 2W 6.001 RESF A 1MZ 0.02% 2W 7.001 RESF A 100KZ 0.02% 2W 8.001 RESF A 10KZ 0.02% 2W 9.001 RESF A 1KZ 0.02% 2W 10.001 OPBR Does UUT have AC function? 10.002 JMPF 10.003 HEAD AC VOLTS 10.004 5100 10 10V 0.15% 50H E 2W 11.001 5100 100 100V 0.15% 50H E 2W 12.001 5205 1000 750V .14% 50H E 2W 13.001 5205 1000 750V 0.14% 10KH 2W 14.001 5100 100 100V 0.15% 10KH E 2W 15.001 5100 10 10V 0.15% 10KH E 2W END